8.3.1.E
Span Calibration Reminder Setup
To get here, navigate:
[MENU]
CALIBRATION MENU
BALANCE CALIBRATION MENU
SPAN CAL REMINDER SETUP
A span calibration reminder prompt is
available to automatically remind you to
run the balance span calibration at the
interval dictated by your policies or
procedures.
The following options are available:
•
ENABLE turns on/off the reminder.
The factory default setting is
No
(off).
•
TIME specifies the time that the
prompt appears using a 24-hour
format. This is usually set to the start
of the shift.
•
INTERVAL - The interval between calibrations in weeks. If the interval is set to “0”, the
reminder will occur daily.
8.3.2 Temperature Calibration
To get here, navigate:
[MENU]
CALIBRATION MENU
TEMPERATURE CALIBRATION MENU
All six menu options shown are available
on the MAX
®
4000XL, but on the MAX
®
4000, only the first option VIEW
CALIBRATION REPORT is available,
since Temperature Calibration and
Verification functionality is only available
on the MAX
®
4000XL.
The MAX
®
4000XL temperature
calibration routines calibrate and verify
the accuracy of the oven temperature
throughout the usable range. The oven is
calibrated at 10 preset increments and
verified at user selectable temperatures.
The Temperature Calibration Kit is required (See section
). This kit contains an NIST traceable device that measures the temperature
at the sample position.
Note: For optimum performance, perform the calibration in an environment that is
free from air currents and general activity (evenings or smaller staffed shifts are
best).