![Arizona Instrument COMPUTRAC MAX 4000 User Manual Download Page 85](http://html1.mh-extra.com/html/arizona-instrument/computrac-max-4000/computrac-max-4000_user-manual_2972205085.webp)
AZI Sales/Customer Service 800-528-7411 or 602-470-1414 or [email protected]
Page 85 of 105
8.7.
Sample Foils
To get here, navigate:
[MENU]
SAMPLE FOILS
This begins the process of weighing and storing foil weights up to a maximum of 50
unique foils. The foils are weighed before the sample is applied. After weighing, the foils
are used in paint processes where the paint is sprayed onto the foil as the end product is
being painted. When placed into the MAX
®
4000, the foil weight is subtracted from both
the start weight and end weight so that only the spray product will be considered in the
moisture determination.
To use the foils for testing, SPECIAL TARE (in the Test Program) must be set to Foil (See
section
).
Adding new foils:
•
Press
[ADD]
. The first foil in the list is
the last one added (LIFO). Note: The
cooling fan must be off before
weighing foils for the list, and there is
no confirmation prompt before
deleting a foil.
•
Follow the prompts to open the lid,
place a foil on the sample pan, and
close the lid.