![Arizona Instrument COMPUTRAC MAX 4000 User Manual Download Page 12](http://html1.mh-extra.com/html/arizona-instrument/computrac-max-4000/computrac-max-4000_user-manual_2972205012.webp)
AZI Sales/Customer Service 800-528-7411 or 602-470-1414 or [email protected]
Page 12 of 105
•
A User Login function that, if activated, can be used to restrict access to calibration
functions and instrument and test configuration options except by authorized users,
preventing unwanted changes to operating parameters. When User Login is active on
the MAX
®
4000XL, a user is required to login prior to testing, and the user name is
recorded with each test result.
•
On 21 CFR, Part 11 compliant instruments, the Delete Test Results and Delete Audit
Log Events functions are disabled throughout the system.
For those familiar with moisture analysis instruments, the
MAX
®
4000 should be simple to operate. However, please read
all of the sections in this manual before attempting your first
analysis.
Contact AZI Customer Service if you have any questions about the following instructions.
An up-to-date electronic copy of this manual can be found on the AZI web site at
5.1.
Hardware
The MAX
®
4000 series instruments consist of:
•
A small oven that dries the sample.
•
A precision balance weighing mechanism with a resolution of 0.001 gram (0.0001
gram on the MAX
®
4000XL) and a full scale range of 40 grams that continuously
weighs the sample.
•
An input power module that can be set to 100-120 VAC or 220-240 VAC.
•
A large backlit color graphic LCD.
•
A 32-bit microprocessor that controls the test process, data handling,
communications, and performs all necessary calculations.
•
A lid fan that rapidly cools the heat chamber to reduce time between tests to increase
overall throughput.
•
A small fan in the lower chassis that maintains the electronics at a stable and low
temperature.
•
Connectors on the back panel include:
o
A USB Type A / Host port for connecting a USB printer, a USB keyboard, a
Nitrogen Control module, a bar code reader, and/or other accessories.
o
An RS-232 serial port for PC connection and Temperature Calibration Interface
(TCI) module connection.
o
An Ethernet connection for using the optional Web Server. (MAX
®
4000XL only).
o
A USB Type B / Slave port for connecting a computer to the MAX
®
4000.
The MAX
®
4000 is designed to remain powered up at all times. From a cold start, the
instrument requires a 60 minute warm up period for optimum accuracy. When the
instrument is not in use, select a test with an idle temperature of 25 ºC and keep the lid
down. The oven will remain cool and the LCD backlight will turn off after a period of
inactivity to conserve energy. Press any key or open or close the lid to restore illumination
to the LCD.