FEATURE
SPECIFICATION
Balance
Integrated balance with automatic operation.
Balance capacity of 40g.
Standalone balance mode available when not testing samples.
Resistant to nominal vibration with options available for
isolation feet.
Balance Span
Calibration
Menu driven calibration and verification by the end user.
Programmable calibration reminder.
Balance Linearity
Verification
2 to 8 points at either a 0.5, 1, 2, 3, 5, and 10g increment and up to
40g total. (MAX
®
4000XL only)
User Interface
Menu driven user programmable interface through keypad and
large color display.
Display
¼ VGA, 320 x 240 pixel color with adjustable contrast via contrast
screen
Electronic Interface
USB host for USB printer, keyboard or USB drive connection.
Serial port for PC connection and Temperature Calibration
Interface (TCI) module connection.
Ethernet for Web Server (Optional, MAX
®
4000XL only)
USB connector for computer connection.
Web Server
(Optional)
(MAX
®
4000XL only)
Web Server used with Microsoft Internet Explorer
®
allows a user
to:
•
Download a complete calibration report.
•
View the LCD screen
•
Download and upload Test Programs.
•
Download Test Results and Test Graph data.
•
View and download the Audit Log
Memory
Storage for:
MAX
®
4000:10 Test Programs (6 user programs and 4 factory),
2 Test Results and no Test Graphs.
MAX
®
4000XL: 250 Test Programs (245 user and 5 factory),
1,000 Test Results with the last 100 Test Graphs
Statistical Calculation Mean, standard deviation, relative standard deviation
21 CFR, Part 11
compliant
instruments
The Delete Test Results and Delete Audit Log functions are
disabled.
User Login
Up to 20 users can be configured with username and passkey
•
When active, a user must be logged in before a test can be run
(MAX
®
4000XL only).
•
Standard user: Can perform tests, view data and reports.
•
Configuration user: Can also change setup configuration,
add/edit/delete Test Programs, and delete data.
•
Calibration user: Can also perform balance and temperature
calibrations.
Self-diagnostics
Built in hardware and software diagnostics software
Lid Fan
Rapidly cools the heat chamber to reduce time between tests to
increase overall throughput.
Options
Temperature calibration interface kit (MAX
®
4000XL only)
Printer kit
IQ/OQ/PQ Validation Package (MAX
®
4000XL only)