![Arizona Instrument COMPUTRAC MAX 4000 User Manual Download Page 41](http://html1.mh-extra.com/html/arizona-instrument/computrac-max-4000/computrac-max-4000_user-manual_2972205041.webp)
AZI Sales/Customer Service 800-528-7411 or 602-470-1414 or [email protected]
Page 41 of 105
8.1.
Test Programs
Test Programs are test parameter sets. They are sequentially numbered from 1 to the
total number of programs entered into the instrument. The figure below shows program
1 selected, as indicated by the yellow arrow. The 1/20 in the upper right hand corner
indicates that a total of 20 test programs have been defined and the blue cursor bar is on
number 1 out of 20. The “(Factory)” programs are for standard tests to check the
performance of the instrument. The MAX
®
4000XL has 245 user programs and 5 factory
programs, while the MAX
®
4000 only has 6 user programs and 4 factory programs. (The
(Factory) DRIFT test is only available on the MAX
®
4000XL.)
To get here, navigate:
[MENU]
TEST PROGRAMS
TEST PROGRAMS allows access to the
test parameter sets. This is where the
contents of the Test Programs may be
printed or changed.
The keypad functions are:
•
Up and Down Arrow Keys
move the
cursor one line at a time.
•
Left and Right Arrow Keys
move the
cursor one page at a time.
•
[START]
key selects the highlighted Test Program and immediately begins a test.
•
[ENT]
– Select and edit the highlighted Test Program. Press
[ESC]
when done to
return to the TEST PROGRAM EDIT screen. This program will now be selected.
•
[CUT]
– Copy the current program into temporary memory and delete it from the list.
The program can be optionally inserted at another location.
Use this carefully.
The
“(Factory)” programs cannot be deleted with this option.
•
[COPY]
– Copy the current program into temporary memory, which can be optionally
inserted at another location.
•
[INSERT
↑
]
or
[INSERT↓]
- Insert the program that is in temporary memory into the list
either above or below the current program. The following programs are moved down,
and multiple copies can be made with multiple insertions.
•
[I/O]
– Displays the
[SAVE]
,
[LOAD]
,
[PRINT]
, and
[MOVE]
soft keys. These soft
keys are used to upload, download or print Test Programs, as shown on the next
page.
To add a new Test Program:
•
Highlight an existing Test Program and press
[COPY]
,
•
Use the arrow keys to scroll to the desired location in the list for the new program,
•
Press one of the insert keys to insert a copy of the Test Program.
•
Edit the inserted Test Program to reflect the desired parameters, as explained in the
following pages.
When the MAX
®
4000 is powered on, it will immediately begin heating to the IDLE
TEMPERATURE of the first Test Program in the list. When shipping from the factory, the
default first Test Program in the list is the (Factory) IDLE Test Program, which has an
IDLE TEMPERATURE of room temperature (25 °C) to avoid undesired heating. To
change the default Test Program upon power up, insert the desired Test Program at
position 1 in the list of Test Programs.