8.3.2.B
Verification Setup
Available on the MAX
®
4000XL only
To get here, navigate:
[MENU]
CALIBRATION MENU
TEMPERATURE CALIBRATION MENU
VERIFICATION SETUP
Verification Setup allows customization of
the temperature verification step that is
performed after the temperature
calibration.
The following options are available:
•
AUTO VERIFY AFTER CAL enables
the auto verification following a
calibration. If time is critical and only
calibration is desired, then disable
auto verification.
•
ADDITIONAL VERIFY TOL – The pass/fail tolerance criteria for the verification can be
increased from the factory calibration standard of
±
1.5 °C to match your quality
standards or standard operating procedures (SOP’s).
•
POINTS TO VERIFY options are:
o
All Cal Points – All 10 standard calibration points are verified.
o
1/2/3/4 Custom Pt(s) – Up to 4 custom calibration points can be verified instead of
the 10 standard calibration points.
•
CUSTOM TEMP VERIFICATION POINTS – If custom calibration points are going to
be used, those temperatures are specified here. Up to 4 custom calibration points can
be verified instead of the 10 standard points, if POINTS TO VERIFY is set accordingly.
8.3.2.C
Calibration Module Setup
Available on the MAX
®
4000XL only
To get here, navigate:
[MENU]
CALIBRATION
MENU
TEMPERATURE
CALIBRATION MENU
CALIBRATION MODULE SETUP
This menu item allows entry of the
Temperature Calibration Kit’s module
number, probe number, and certification
date. This information is located on the
Temperature Calibration Interface (TCI)
Module.