
6
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
-
If the pump shall be operated at very low rates,
arcomed ag
recommends the use of the
Syramed® µSP6000 Chroma syringe pump.
-
Warning: The Volumed® µVP7000 Chroma must not be used for patients, which are
generally contra indicated to intravenous application of medicine.
-
It is the responsibility of nursing staff to station a replacement infusion pump near the
patient if a temporary failure of the pump would create a not acceptable risk for the
patient.
-
Check the pump and accompanying accessories for damage when unpacking. The pump
must not be operated if damaged. Should the pump be damaged contact your
arcomed-
distributor.
-
The Volumed® µVP7000 Chroma is equipped by the manufacturer with the following
accessories: empty Bag detector, user manual, mains cable.
-
Autoclaving the pump is not allowed.
Fogging is permitted. Make sure there’s no moist
before taking the pump into operation.
-
The user must always verify that the set rate does not exceed the maximum flow
capability of the catheter or the canola.
-
The correct function of the pump can only be guaranteed if the pump is used with the
specified infusions devices. If you want to use any other infusion lines than specified on
top of the pump, please get into contact with your
arcomed
distributor.
-
The Volumed® µVP7000 Chroma should normally be operated from a mains power
supply. The internal batteries will automatically operate the pump in the event of a power
failure.
-
Warning:
The patient is not allowed to be in a water tub, take a bath or similar
whilst connected to the pump.
-
Do not operate this pump in an environment with high levels of electromagnetic radiation
such as MRT unless used with the Arcomed Unique MRI Shield. Refer to the instructions
and setup guidelines of the Unique MRI Shield manual.. The properties of the pump may
be influenced causing a hazardous situation. Mobile telephones may never be used
anywhere near the pump. For additional information also see chapter 9.
-
This pump is not designed for use in areas where there is an explosion hazard.
Environmental requirements as per IEC601-1-2 must be observed. For further
information contact the official distributor in your country or the Customer Service
Department in Switzerland:
arcomed ag
, Althardstrasse 150, CH-8105 Regensdorf, Schweiz
Tel. +41 (0)43 388 90 30, Fax. +41 (0)43 388 90 40
-
Warning: To avoid a Free Flow situation, always close the roller clamp when
changing infusion devices.
Summary of Contents for Volumed uVP7000 Premium
Page 1: ...Edition 03 2013 ...