14
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
Figure 4: Stopf Flow test in progress
NOTE:
During the ON/OFF cycle the pump emits an audible alarm and all alarm
indicators, all status indicators and all LED displays (8888) are illuminated. Additionally
the LCD displays the actual SW Version. If the audible sound is not emitted or any
displays are defective, the pumps must not be used.
If the test was passed the menu can be displayed by pressing the SELECT key.
Choose the option „
Prime/Bolus
“ by pressing the DATA keys.
Figure 5: Overview of functions
After confirming with SELECT the infusion line can be purged by pressing and
holding the II key on the panel. Note: The touch screen can not be used to prime the
line for safety reasons.
Summary of Contents for Volumed uVP7000 Premium
Page 1: ...Edition 03 2013 ...