28
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
At next the maximum amount of drug that can be delivered in a certain time is entered:
Figure 29: Entering the amount of drug (20.00 mg / 0.10 h / 20.00 mg / 10min)
Now the constant rate is set:
Figure 30: Entering the continuous rate (4.0 mg/h, which equals with the set concentration 4.0 ml/h)
If the constant rate shall only run for a certain time, the continuous time can be set:
Figure 31: Entering the continuous time (15 min)
When all parameters are entered the infusion giving set can be loaded. After that, the roller
clamp can be opened and the pump can be started
by pressing the “Start/Stop” key.