The Instructions for Use, the mains power supply cable and the empty bag detector are
included as standard equipment with the Volumed® µVP7000 Chroma.
3.11
Using the Pump in Parallel or Multiple Infusions
If additional infusion systems are connected to the patient's vascular system, this may lead to
complications e.g. infusion of air, reverse-flow, interruptions due to alarms and inaccurate
flow.
To prevent such incidents, please observe the recommendations as stipulated in DIN VDE
0753, Part 5 or contact your distributor.
4
The Alarm System
4.1
Alarm Indicators
The alarms conditions are indicated using illuminated pictograms as shown below:
4.2
Alarm Causes
The electronic self-monitoring system continuously monitors the correct function of the pump
and displays system whilst the pump is in operation. If a fault condition should occur, the
infusion stops immediately and the alarm is activated. An appropriate alarm symbol (red
coloured) is illuminated continuously, accompanied with a continuous audible alarm. The
nurse call alarm is activated at the same time.
By pressing the “Alarm Silence” key (SELECT) the acoustic alarm can be suppressed for 2
minutes:
The Volumed® µVP7000 Chroma will not start:
If the infusion set is not inserted correctly in the air in line detector and the stop flow
mechanism,
if no rate has been entered (0.0 ml/h),
if the door is opened,
if the Robson clamp is not inserted correctly (Stop Flow)
The pump only switches to KVO rate during infusion accompanied with an audible alarm
when:
the “Start/Stop” key is pressed,
the pump has infused the VTBI set (infusion complete),