Guidance and manufacturer's declaration - electromagnetic emission
The Volumed µVP7000 Series is intended for use in the electromagnetic environment specified below. The customer or
the user of the Volumed µVP7000 Series should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Volumed µVP7000 Series uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Volumed µVP7000 Series is suitable for use in all establishments
and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class B
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
The Volumed µVP7000 Series is intended for use in the electromagnetic environment specified below. The customer or
the user of the Volumed µVP7000 Series should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 610004-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power quality should be that of typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV
differential mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality should be that of typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T
)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in
UT)
for 5 sec
<5 %
U
T
(>95 % dip in
U
T
)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
As the Volumed µVP7000 Series is powered by an
internal battery, no special precautions must be
taken.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
400 A/m
400 A/m
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.