the tubing is in a straight line. Put the Robson clamp with your thumb into the Stop
Flow device (B) and insert the remaining tubing onto the right tube guide (A).
Slot the drip chamber into the empty container detector. Make sure that there are no
large ribs or joints in the passage of the empty container detector and that fallen
drops are detected by the drop-
detector‘s light-barrier.
Figure 2: Pump with PVC infusion set
Figure 3: Pump with silicon infusion set
Warning: Pumps calibrated to a silicon infusion set may only be used with a infusion
set with silicon segment. Always insert the silicon segment in such pumps and not the
PVC tubing (consider picture above). The connectors, where the silicone is attached
to the PVC must be left and right of the tube guides (C)!
Close the door (12) and push the latch (13) firmly against the pump. Open the tubing
roller clamp.
Now the Stop Flow test is performed