9
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
3
Operation
3.1
General Remarks
The figures in brackets refer to the illustrations of front and rear views shown on pages 2 and
3.
Warning: Use only approved infusion sets.
The performance of the pump depends on both pump and administration set. The Volumed®
µVP7000 Chroma must only be used with the sets the pump has been calibrated to.
The functional safety of the pump cannot be guaranteed if non-approved infusion sets are
used. The safety of the patient may be compromised as a result.
Disposable IV sets are for single-use only. Single-use needles carry an infection hazard and
must be disposed of in accordance with local guidelines. PVC infusion sets should be
replaced every 24 hours, silicon infusion sets after 72 hours.
Be aware, that the audible alarm interval and initial pitch and sound level can be configured
to the customers’ needs. Check if the current setting of the audible alarm is appropriated to
the environment before connecting the pump to the patient.
The Volumed pump can be configured to the customers needs, please contact the arcomed
representative for detailed information on you pumps configuration settings.
Very important:
The infusion set brand to which the pump is calibrated is indicated by a sticker on top of the
pump.
3.1.1
Basic concept of navigation:
The Volumed® µVP7000 Chroma is equipped with a large high definition color LCD display.
A bottom line splits the LCD in two parts (see Figure 1). The upper part contains icons of
functions and modes. The lower part contains the SELECT button to confirm, the ESC button
to get back and if available several scrolling arrows to the left or right might be available. This
allows scrolling to the previous or next page of selection.
Figure 1: Structur of LCD display
Upper Part
Bottom line
Lower Part