concentration is changed, the pump will react by applying a bolus or it will stop for a while
before the new permanent rate is activated.
While the pump is infusing, the rate in ml/h is displayed in the rate window. The target
concentration is displayed in the DATA window. The LCD shows the target and the actual
concentration. To change the target concentration, simply adjust the value in the DATA
window and confirm with the SELECT key.
Figure 40: Pump in TCI mode (Rate = 47.3 ml/h (RATE window), cet = 3.00
g/ml (DATA window), ce = 3.00
g/ml (LCD), rate in unit: 13.5 mg/kg/h, total infused volume 89.4 mg, remaining time until syringe is depleted: 29
min 48 sec (LCD)). This lower display alters with the patient data (weight, height, age). The short cut Awake conc.
(optional) allows to change the awake concentration, the syringe short cut will show all infusion parameters.
Besides the functions described in chapter 3.4 and 3.5 the following TCI specific functions
are available by pressing ESC:
Figure 41: LCD display, called by pressing ESC while the pump is running.
Figure 41a: Graphic displays of the actual and future concentration. The vertical line symbolises the actual state,
left is passed, right is future. The hatched area shows ce, the horizontal line shows cet, the rising and falling line
shows cp.
Figure 41b: Info for wake up time, total infused amount, rate in mg/kg/h and total time (alternating).
Figure 41a
Figure 41b