1.2
General Information
The Volumed® µVP7000 Chroma is a volumetric infusion pump for medical applications. To
assure highest standards in quality and user-friendliness latest technology in R&D and
production are used.
This microprocessor controlled volumetric infusion pump operates by pumping the infusate
using a peristaltic mechanism. The sterility of the infusate is not affected. The pump is
designed to infuse drugs or other infusates into the patient by controlled means under
pressure.
The pump can be used for infusing small and large volumes. The rate (0,01 ml/h - 1200 ml/h)
and the Total volume are precisely adjustable and reproducible.
The pump is driven by a stepper motor. A drive belt transfers the movement to a peristaltic
mechanism. The peristaltic fingers squeeze the tubing in one direction, transporting the
liquid. All infusion parameters are visualized on LCD and LED displays. Input of the desired
values is done by a key pad.
The Volumed® µVP7000 Chroma can be used in both stationary and transportable
applications due to a long battery life span up to 6 hours of duration. Applications include
Neonatology, Intensive and Cardiac Care, Paediatrics, Gynaecology and Obstetrics,
Surgery, Pain Treatment, Anaesthesia and General Medicine. It can also be used in
Ambulances and Air rescue.
Warning: It is forbidden to use Volumed® µVP7000 Chroma for enteral application.
To be informed about technical changes and incidents related to this device; please consult
the
arcomed
The Volumed® µVP7000 Chroma may be operated only on mains power installed to DIN
VDE 0100-710:2002-11 or the appropriate national standards.
The Volumed® µVP7000 Chroma is a medical device and meets the Medical Device
Directive (MDD) requirements of the EC Guideline 93/42 EEC and is marked CE 0123 (TUV
Süd PS Munich, Germany).
Manufacturer according to MDD is
ag
, Althardstrasse 150, CH - 8105 Regensdorf,
Switzerland.
Classification according to MDD: IIb
1.3
General Information, Contraindications
The Volumed® µVP7000 Chroma may only be used by trained staff. Attention should be paid
to all security measures, installation notes and issues as stated in the user manual:
-
Using infusion sets or connecting accessories that have not been tested and released by
arcomed ag
for the use with this pump can harm patients, users and other people and
therefore is not allowed.