21
GA 7000C-E-0313
© 2013 by arcomed ag, Switzerland
2.
Setting the drug concentration by pressing the DATA keys, to confirm press SELECT
key.
Figure 16: Entering the drug concentration (5.00 mg/ml)
3.
The rate is set in the corresponding unit in the DATA window. The rate window
displays the calculated rate in ml/h.
Figure 17: Setting the rate (10.0 mg/h), this corresponds to 2.0 ml/h
4.
The rate in ml/h can always be checked by using the following formula:
ml
mg
ion
Concentrat
h
mg
Rate
h
ml
Rate
5.
The pump is started with the „START/STOP“ key.
Summary of Contents for Volumed uVP7000 Premium
Page 1: ...Edition 03 2013 ...