The infusion time is entered in the rate window, the VTBI in the DATA window. As data is
entered it is simultaneously displayed on the LCD.
Once all values are entered, the
„START/STOP” key may be pressed to start the infusion.
NOTE:
9 hours, 30 minutes are entered as 9.30 in the rate window.
Figure 10: Input screen for infusion time (rate window) and VTBI (DATA window).
3.3.4
Choosing a Drug from the Drug Library
HINTS:
The Volumed® µVP7000 Chroma has almost unlimited possibilities of configurations. As a
pump a
dministrator it’s highly recommend to
KEEP IT SIMPLE
. Many nurses are still used to
simple ml/h infusions and before getting into complicated drug schemes make sure the staff
is trained to master and most importantly not confuses e.g. drug dose calculation or TCI with
classic ml/h infusions.
The Volumed® µVP7000 Chroma also allows to have various configurations, typically a
configuration for general ward, ICU or OR. With adequate startup drugs and short cuts the
risk of wrong entries (drug dose errors) can be significantly reduced. Take advantage of
these features where ever possible.