3.3.2
Individual Program - Set up of Rate and VTBI
To use the device in the simple individual mode, after priming rate in ml/h and VTBI can be
entered by pressing the rate and DATA keys. Use arrow up keys to increase rate or VTBI,
arrow down keys to decrease rate or VTBI. Note each key press produces a single digit
change. The least significant digit (smallest size) indicated is 0.1 ml/h if the rate does not
exceed 999.9 ml/h, otherwise the least significant digit indicated is 1 ml/h.
Figure 8: Pump with rate 41.6 ml/h and VTBI 1000.0 ml
NOTE:
To recognize a possible malfunction of display elements, it is necessary to verify that
the displayed rate is changing in accordance with the input sequence.
In case that only a certain amount shall be infused, the DATA keys can be used to set the
desired volume in ml. The infusion is started by pressing the
„START/STOP“ key.
If the pump is used together with the empty bottle detector (Easy Clip) the pump can be
started without entering a VTBI. Is the empty bag detector not installed a target volume
(VTBI) must be entered before the pump can be started. The entered VTBI must not exceed
the actual volume in the bag. If a VTBI is entered, the pump will show it below the infused ml.
3.3.3
Time over Volume Function
This function allows infusing a specific volume in a certain time. The calculation of the rate is
done by the pump. To enter this function press SELECT and scroll by pressing the DATA
keys
to the menu point „
TOV
“. To confirm, press SELECT. If the touch screen is enabled,
you can directly press on the TOV icon without scrolling.