Furthermore the Volumed® µVP7000 Chroma has a very powerfull interface combined with
the Unique Concept. This allows not only to monitor the pumps and record the events but
also to update automatically patient data or the latest drug mode libraries.
Drug mode libraries allow nursing staff to deliver a certain drug without any rate calculation in
ml/h. The drug library must be programmed into the pump previously. Choosing a drug will
also load the drug specific infusion parameters like VTBI, rate, hardlimits and softlimits.
To run the pump in this mode SELECT is pressed after the infusion set was loaded. Press
the DATA
keys to scroll to the menu point „
Drug Classic ml/h
“. If the touch screen is enabled,
you can directly press on the Drug Classic ml/h icon without scrolling.
Figure 11: Menu for selecting the classic drug library.
By pressing SELECT again, the mode is started. The saved drugs can be chosen with the
DATA keys and are confirmed with the SELECT key.
The pump can also be set up to have drug classes or folders containing a sub set of drug
mode selections. This allows having a better overview if the drug mode library contains many
drug modes. To access a folder select it by pressing on the icon or scrolling until the folder is
selected (yellow text underlay) and press SELECT. To exit a folder or subclass continue
scrolling passing the last or first drug in the class. To get back to the main window press
ESC. This aborts the drug selection.