A.3
Command Summary
A.3.1
Battery Charging Status (BAT)
Command:
BAT
Returns Battery and charging status.
Return Value
Meaning
0
Nothing
1
Low Battery
2
Charging
A.3.2
Backlight Set (BLT)
Command:
#BLT
Controls the front panel LCD backlight.
#
Response
0
Backlight OFF
1
Backlight ON
2
Backlight AUTO. The backlight remains on
for approximately 20 seconds after the last
keystroke.
A.3.3
Channel Calibrate (CCH)
Command:
ch#,ac/dc#,range#,gain#,ph/off#,s/r#CCH
Sets or returns the calibration constants for Channel 1 and Channel 2.
Each channel has a set of calibration constants for each range.
#
Action
1
Select Channel 1
2
Select Channel 2
ac/dc#
0
DC
1
AC
range#
0 . . . 8
(see SRG command (below)
gain#
Corrected gain = uncorrected gain
×
(1 +
Gain#
÷
1000)
ph/off#
Corrected Phase = uncorrected phase +
(ph/off#
×
frequency)
Offset = ph/off#
÷
1000 (DC)
107