1,1997,1,1,0,0,0SL0: Sets the autolog stop time to January 1, 1997 at
midnight. Stop time must be later than start time.
The configuration of custom data is contained in a 32-bit word. Each
bit in this word refers to a data parameter. The custom data Table B.1
lists each of the available parameters along with their assigned ID bit and
decimal weight. To remotely program the custom word add the decimal
weight of each desired parameter. The following example shows how to
compute the custom data word for logging time and channels 1 and 2
magnitude data.
Parameter
Decimal weight
Time
1
Channel 1 rms
512
Channel 2 rms
32768
Custom configuration value
33281
B.3.7
Review of Logged Data
#1,#2RLD selects a log record for review or removal.
If #1 and #2 are absent: A list of stored records is returned.
If only 1 number is present:
#=0
All records are cleared including flash memory if
present
#
>
1
The specified record data overwrites current in-
strument data and measure hold is enabled to
allow data review or return.
If both numbers are present:
#1=0
Clears file selected by #2
#1
>
0
Reviews the record indicated by #2 of the file
indicated by #1.
137