A.3.36
Transducer Channel (SCC)
Command:
ref#,r1#,r2#,s/r#SCC
Includes DC Reference, and Shunt Resistance Values – Set or Return (see
Note 1 on page 105). Sets or returns the transducer channel DC reference
and shunt resistance values as follows:
ref#
Reference voltage in volts
r1#
9 ohm shunt resistance in ohms. This calibrates the
1 to 100 mA DC current ranges.
r2#
1000 ohm plus 9 ohm shunt resistance in ohms. This
calibrates the 0 to 1 mA DC current ranges.
s/r#
0
To store reference and resistance values.
1
To send reference and resistance values.
A.3.37
Serial Number (SER)
Command:
serial#,s/r#SER
Sets or returns instrument Serial Number. See Note 1 on page 105.
Serial# (s/r#)
ASCII number (5 digits)
0
To enter serial number
1
To return serial number
A.3.38
Flash Log (SLO)
Command:
#1,#2,#3,#4,#5,#6SLO
Sets the Flash Log operation. See Appendix B for use of this command
with the Extended Memory Option.
1. If no numbers present, it starts a new file
2. If 1 number present: 0 = disable flash, 1 = enable flash
3. If 3 numbers present:
#1 = 0
manual trigger from LOG DATA key or RS-232
Log command
#1 = 1
trigger from timer
#1 = 2
auto trigger with period set by #2
#2 = 1
to 86400 auto log store trigger period in sec
#3 =
32 bit data descriptor entered as decimal number
4. If 7 numbers present
#1 = 0
#2 . . . #7 = autolog start time
#1 = 1
#2 . . . #7 = autolog stop time
#2 =
yyyy year
118
Summary of Contents for 931A
Page 6: ...vi ...
Page 12: ...xii ...
Page 14: ...xiv ...
Page 16: ...xvi ...
Page 52: ...SOURCE LOAD B A C N Figure 3 12 3 Phase 4 Wire 3 Element Connections 36 ...
Page 90: ...Iout I1 V1 V2 V3 I2 I3 Figure 3 17 3 Phase 4 Wire 3 Element Transducer Test 74 ...
Page 114: ...98 ...
Page 120: ...104 ...
Page 142: ...126 ...