Action
0
(or # missing) – return frequency immediately
1
Return only if there is a new reading
A.3.14
Timer Calibrate (FTM)
Command:
c#,s/r#FTM
Sets or returns the calibration constant to correct the timer time and
frequency reference. 1 + (c#
÷
1,000,000) multiply the time measurements
made by the timer.
c#
Timer Calibration Constant
s/r#
Action
0
Set Timer Calibration
1
Return Timer Calibration
A.3.15
Harmonics Display (HAR)
Command:
harm#,ch#HAR
Displays the selected harmonic (with others) for the selected channel.
Shows magnitude, phase and frequency data for the harmonic number
specified as harm#.
Ch#
To select
1
Channel 1
2
Channel 2
A.3.16
Harmonic Magnitude, Return (HMA)
Command:
harm#,ch#HMA
Returns individual harmonic magnitudes for the selected channel.
Set
the unit to measure harmonics, on the selected channel using the HAR
command, prior to issuing the HMA command.
harm#
To select
2–50
Harmonic number (2 through 50)
1
THD (Total harmonic distortion)
ch#
1
Channel 1
2
Channel 2
Response: Numeric ASCII string.