Transducer, continued
Range:
±
10 Vdc
±
100 mAdc
Input
Impedance:
100 kΩ (voltage)
Burden:
20 Ω nominal, including PTC resettable fuse
1 kΩ for
≦
1 mA
4.1.4
Timer
Inputs:
Two
Maximum:
300 Vrms or DC
Isolation:
300 Vrms
Threshold: Contact open
2.9 Volts
Threshold: Contact closed
2.6 Volts
Source Impedance
≈
25 kΩ
(for dry contact):
4.2
Measurements
4.2.1
Voltage and Current
Method:
Wideband: True RMS, 3 kHz BW
Narrowband: Fundamental Magnitude
Accuracy:
0.05% of reading
Under range:
<
1% of reading, typical at 0.3
mArms
DC Voltage: 0.1% of r 25 mV (typical)
4.2.2
Phase Angle
Input:
Channel 1 to Channel 2
Range:
0 – 360
◦
or
±
180
◦
Accuracy:
0.05
◦
Under range:
<
1 degree, typical at 0.3 mArms
100