6.
Display Output As
– select either Measured Quantity or Percent
Error. Although the menu choices include Percent Full Scale, only
Measured Quantity and Percent Error pertain to Custom Pulse.
The sample Transducer Configuration Menu shown below illustrates
settings for a different transducer using the “Rate of A” Function instead
of “Time A to X Counts.” Notice the large difference in Output Range
compared to the “Time A to X Counts” configuration above.
When all of the six categories are set up correctly, press ESCAPE to
return to the main menu. If initially setting up a transducer custom pulse
test, proceed to the section on Timer Configuration below.
3.8.5.4
Configuring in the Timer Menu
Press the Timer CONFIG key to access the Timer Menu. A sample menu
is shown below.
Only two selections under Timer Function apply to the Custom Pulse
feature: “Time A For X Counts” and “Rate of A.”
Make the Timer Configuration selections as follows:
1.
Function
– select either Time A For X Counts, or Rate Of A. Use
Time A for X counts if the contact closure rate is
<
1/sec. Use Rate
of A if the contact closure rate is
>
1/sec. For more details on which
function to choose, see below under “Contact Closures
<
1/second”,
or “Contact Closures
>
1/second.”
77