3.3.2.12
Config (Transducer)
TRANSDUCER
CONFIG
Press SHIFT then the CONFIG key to set the transducer input
and output ranges; it affects the testing of transducers. The
Model 931A can be set to automatically calculate and display
transducer error as percent of full scale, when the same input
signal is applied to the transducer under test and the Model 931A. Any
calibrator or load box with source may be used, or transducers may be
calibrated in-circuit.
Press the CONFIG key to view the following menu:
Use the arrow keys to select a menu item, then press ENTER. Selections
made from this menu will activate other sub-menus, which allow for various
transducer-test configurations. Once all parameters have been set, press
ESCAPE and Transducer Operation will begin. See Section 3.8 for further
information on transducer testing.
3.3.2.13
Help
MENU
HELP
Press SHIFT then the HELP key to find reference information
about the functions of the Model 931A. To get help for a specific
function, press SHIFT, then HELP, followed by the key for the
desired function.
Use the up and down arrow keys to page
through the help message; press the ESCAPE key to return to normal
operation. Display a menu featuring help topics for general operation by
pressing SHIFT then HELP twice (i.e. SHIFT/HELP/SHIFT/HELP).
3.3.2.14
Review
LOG DATA
REVIEW
Press SHIFT then REVIEW to view a menu of stored records.
An example display is shown below:
38
Summary of Contents for 931A
Page 6: ...vi ...
Page 12: ...xii ...
Page 14: ...xiv ...
Page 16: ...xvi ...
Page 52: ...SOURCE LOAD B A C N Figure 3 12 3 Phase 4 Wire 3 Element Connections 36 ...
Page 90: ...Iout I1 V1 V2 V3 I2 I3 Figure 3 17 3 Phase 4 Wire 3 Element Transducer Test 74 ...
Page 114: ...98 ...
Page 120: ...104 ...
Page 142: ...126 ...