2.4.6
User Options
User Method List
After switching on the instrument, a scroll list of all available methods is automatically shown in the
display. To shorten this list according to the requirements of the user, it is possible to create a user
defined scroll list.
The programme structure requires that this list must have at least one active (switched on) method.
For this reason, it is necessary to activate first all require methods and then to switch off the
automatically activated one if this method is not required.
User Method List, Adaptation
Press [MODE], [Shift] + [6] [0] keys.
Confirm with [ ] key.
The display shows:
Start with [ ] key.
The complete method list is displayed:
Method with a point [•] behind the method number will be displayed in the
method selection list. Methods without a point will not be displayed in the
method selection list.
Press arrow key [ ] or [ ] to select the required method from the displayed
list.
Switch with [F2] key between ‘active’ [•] and ‘inactive’ [ ].
Select next method, activate or inactivate as required and repeat.
Confirm with [ ] key.
Cancel without storing by pressing [ESC] key.
Recommendation:
If only a few methods are required, it is recommended to perform Mode 62 first,
followed by Mode 60.
All user Polynomials (1-25) and Concentrations (1-10) are displayed in the method list, although they
are not programmed by the user. Non-programmed user methods cannot be activated.
<Method list>
selected: •
Toggle: F2
save:
cancel: ESC
<Method list>
>>
20
•Acid demand
35•Alkalinity-p T
30•Alkalinity-tot T
….
Summary of Contents for WSL25 Pro
Page 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Page 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Page 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Page 41: ...Section 5 HydroTest Photometer HT1000 ...
Page 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 141: ......
Page 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Page 152: ...NOTES ...