Prepare the first standard and press [TEST] key.
The display shows the input value and measured absorption value.
Confirm with [ ] key.
The display shows:
Enter the concentration of the second standard
e.g. [Shift] + [0] [.][1]
Press [ESC] to delete a digit.
Press [F1] key to reset numerical input.
Confirm with [ ] key.
The display shows:
Prepare the second standard and press [TEST] key.
The display shows the input value and measured absorption value.
Confirm with [ ] key.
The display shows:
Note:
- Perform as described above to measure further standards.
- The minimum number of measured standards is 2.
- The maximum number of measured standards is 14 (S1 to S14).
If all required standards or the maximum value of 14 standards are
measured, press [Store] key.
The display shows:
The instrument goes back to the mode menu automatically.
Now the concentration is stored in the instrument and can be recalled by
entering its method number or selecting it from the displayed method list.
TIP:
Save all your concentration data in a written form because, in case of power outage (e.g.
changing the battery), all concentration data will lost and must be entered again.
<User concentr.>
S1: 0.05 mg/l
mAbs: 12
<User concentr.>
S1 accepted
S2: + _____
│ ESC │F1
<User concentr.>
S2: 0.10 mg/l
prepare Test
press TEST
<User concentr.>
S1: 0.10 mg/l
mAbs: 150
<User concentr.>
S2 accepted
S3: + ______
│ ESC │ F1 │ Store
<User concentr.>
stored!
Summary of Contents for WSL25 Pro
Page 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Page 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Page 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Page 41: ...Section 5 HydroTest Photometer HT1000 ...
Page 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 141: ......
Page 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Page 152: ...NOTES ...