You might want to use mode 67 to transfer all concentration data to a PC.
User Polynomials
It is possible to enter and store up to 25 User Polynomials.
The programme allows the user to apply a Polynomial up to the 5
th
degree:
y = A + Bx + Cx
2
+ Dx
3
+ Ex
4
+ Fx
5
If only a Polynomial of a lower degree is necessary, the other coefficients are specified as zero (0),
e.g. for the 2
nd
degree is D, E, F = 0.
The values of the coefficients A, B, C, D, E, F must be entered in an academic notation with maximal
6 decimal places, e.g. 12,135,673 = 1,213,567E+02
Press [MODE], [Shift] + [6] [5] keys.
Confirm with [ ] key.
The display shows:
Enter a method number in the range from 800 to 824
e.g. [Shift] + [8] [0] [0]
Confirm with [ ] key.
Note:
If the entered number has already been used to save a concentration,
the display shows the query:
Or
Press [Shift] + [0] or [ESC] keys to go back to method no. query.
Press [Shift] + [1] keys to start entry mode.
The display shows:
Enter the required wavelength
e.g. [Shift] + [2] for 560 nm.
The display shows:
Press arrow key [ ] or [ ] to change between plus and minus sign.
<User polynoms.>
choose no.: ______
(800-824)
<User polynoms.>
Overwrite polynom?
YES: 1, NO: 0
<User polynoms.>
wavelength:
1: 530 nm 4: 430 nm
2: 560 nm 5: 580 nm
3:610 nm 6: 660 nm
<User polynoms.>
y = A+Bx+Cx2+Dx3+
Ex4+Fx5
A: + __________
Summary of Contents for WSL25 Pro
Page 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Page 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Page 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Page 41: ...Section 5 HydroTest Photometer HT1000 ...
Page 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 141: ......
Page 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Page 152: ...NOTES ...