User Method List, Switch All Methods On
This mode function activates all methods. After switching on the instrument a scroll list of all
available methods is automatically shown in the display.
Press [MODE], [Shift] + [6] [1] keys.
Confirm with [ ] key.
The display shows:
Press [Shift] + [1] keys to display all methods in the method selection list.
Press [Shift] + [0] keys to keep the valid method selection.
The instrument goes back to mode menu automatically.
User Method List, Switch All Methods Off
The programme structure requires that the method list must have at least one active (switched on)
method. For this reason, the instrument activates one method automatically.
Press [MODE], [Shift] + [6] [2] keys.
Confirm with [ ] key.
The display shows:
Press [Shift] + [1] keys to display only one method in the method selection
list.
Press [Shift] + [0] keys to keep the valid method selection.
The instrument goes back to mode menu automatically.
User Concentration Methods
It is possible to enter and store up to 10 User Concentration Methods.
Therefore, you need 2 to 14 standards of know concentration and one blank (deionised water or
reagent blank value). The standards should be measured with increasing concentrations and from
the brightest to the darkest colouration.
The measuring range for ‘underrange’ and ‘overrrange’ is defined with -2600 mAbs* and +2600
mAbs*. After selection of a method, the concentration of the lowest and highest used standard is
displayed as measuring ranger. The operation range should be within this range to achieve best
results.
* 1000 mAbs = 1 Abs = 1 E (displayed)
<Mlist all on>
switch on all methods
YES: 1, NO: 0
<Mlist all on>
switch off all methods
YES: 1, NO: 0
Summary of Contents for WSL25 Pro
Page 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Page 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Page 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Page 41: ...Section 5 HydroTest Photometer HT1000 ...
Page 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 141: ......
Page 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Page 152: ...NOTES ...