8.3 2 Point Calibration:
8.3.1 Ensure that 2 point calibration is selected in the setup menu.
8.3.2 Press CAL key, the meter shows “100/CAL1”.
8.3.3 Press ▲ or ▼ key until the display shows “0/CAL1”.
8.3.4 Immerse the dissolved oxygen probe into the saturated anhydrous sodium sulfite
solution (zero oxygen solution). Stir the probe gently.
8.3.5 Press ENTER key to confirm, “Calibration” indicator begins flashing.
8.3.6 Wait for the measured value to stabilise, the display shows “100/CAL2”. The meter
prompts you to continue with second point calibration.
8.3.7 Immerse the dissolved oxygen probe in air-saturated water for 3 to 5 minutes, press
ENTER key to confirm. Wait for the measured value to stabilise, the display shows “END”.
Once calibration is complete, the meter returns to measurement mode automatically.
N.B. Performing a percentage saturation calibration will simultaneously calibrate the
corresponding mg/L (or ppm) concentration value. Therefore, additional mg/L calibration isn’t
required in most circumstances.