3.4 Conductivity/TDS/Resistivity Measurement
1. Press the MODE key until the display shows the
(Conductivity) or
(TDS) or
(Resistivity) indicator.
2. Rinse the conductivity electrode thoroughly with distilled water.
3. Immerse the electrode in the sample solution, and stir the solution gently.
4. Wait for the reading to stabilise, and record the value shown on the display.
3.5 Salinity Measurement
The HC1000 has two salinity measurement modes: practical salinity (unit: PSU) and natural
seawater (unit: ppt).
1. Press the MODE key until the display shows the
indicator and measurement unit
“PSU”, the meter enters practical salinity measurement mode.
2. If necessary, press the MODE key again, the display shows the
indicator and
measurement unit “ppt”, the meter enters seawater measurement mode.
3. Select the desired measurement mode. Immerse the electrode in the sample solution. Stir
the solution gently.
4. Wait for the reading to stabilise, and record the value shown on the display.
3.6 Dissolved Oxygen Measurement
The HC1000 is suitable for measuring dissolved oxygen in water, wastewater, brine and
other liquids. If the intended application is to measure seawater or other water containing
large amounts of salt, please set the salinity coefficient before use.
Some gases and vapours such as chloride, sulfur dioxide, hydrogen sulfide, ammonia,
carbon dioxide and iodine can permeate the membrane via diffusion, and as such their
presence will influence the measurement of dissolved oxygen.
Clean samples are preferred. If the sample contains solvent, grease, sulfide and algae, the
membrane on the probe may be blocked, damaged or eroded.
Summary of Contents for WSL25 Pro
Page 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Page 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Page 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Page 41: ...Section 5 HydroTest Photometer HT1000 ...
Page 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 141: ......
Page 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Page 152: ...NOTES ...