2.3.7
Performing Tests
When zero calibration is complete, remove the vial from the sample chamber and perform the tests
as described in the relevant application note.
When the results have been displayed, options include:
-
Storing and/or printing the results
-
For some methods, changing between different chemical species
-
Performing further analysis with the same zero
-
Selecting a new method
2.3.8
Changing Chemical Species
For some methods, there it is possible to change the chemical species of the test results. When the
test result is displayed, press the arrow key [ ] or [ ].
Example:
320 Phosphate LR T
---- [ ] ----> 320 Phosphate LR T <---- [ ] ---- 320 Phosphate LR T
0.05-4 mg/l PO
4
0.02-1.3 mg/l P
0.04-3 mg/l P
2
O
5
1.00 mg/l PO
4
---- [ ] ---->
0.33 mg/l P
<---- [ ] ----
0.75 mg/l P
2
O
5
If the species of a test result is changed, the displayed range is adjusted automatically. For an already
stored result, it is not possible to change the chemical species. The last displayed chemical species is
kept by the instrument and will be displayed if this method is used again. If there is the possibility to
change the chemical species for a method, it is described in the application note. The arrows
indicate the possible chemical species and are printed below the application note.
PO
4
P
P
2
O
5
Summary of Contents for WSL25 Pro
Page 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Page 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Page 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Page 41: ...Section 5 HydroTest Photometer HT1000 ...
Page 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 141: ......
Page 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Page 152: ...NOTES ...