![Aquasafe WSL25 Pro Instruction Manual Download Page 40](http://html.mh-extra.com/html/aquasafe/wsl25-pro/wsl25-pro_instruction-manual_2960690040.webp)
SECTION 9: SELECTING THE OPTIMUM VOLUMES FOR MEMBRANE
FILTRATION
The optimum volume of sample is that which will allow the most accurate quantification of
bacterial colonies. This is achieved when the number of faecal (thermotolerant) coliform
colonies on the membrane following incubation is between 20 and 200 colonies. If there are
fewer than 10 colonies, then there exists the possibility of statistical error. Numbers greater than
200 colonies are difficult to count with the naked eye.
Potable Waters
The number of faecal coliform bacterial colonies in treated water samples should ideally be
zero. Thus, the preferred sample volume is 100 mL, and a count of zero faecal coliform bacteria
per 100 mL is indicative of a microbiologically safe water supply. If the count exceeds 1 faecal
coliform per 100 mL, contamination is indicated. If the count exceeds 10 faecal coliforms per
100 mL, action is urgently required.
Raw Waters
For source waters and partially treated waters, including those which are ground water derived,
it can be useful to adjust the sample volume in order to obtain faecal coliform counts in the
optimum range 10-200. It may also be useful to process more than one quantity on the first
occasion a particular water source is sampled. In such cases it is not necessary to re-sterilise
the filtration equipment between different quantities of the same sample, provided that the
smaller volume is processed first. Typical volumes which may be appropriate for various water
types are shown in the following table. They are only guidelines; there is no substitute for
experience of a given source.
APPROPRIATE
VOLUME (mL)
SOURCE OF SAMPLE
100
50
10
Lakes, Reservoirs, & Rivers &
other surface sources
*
**
***
Wells, boreholes, other protected
water sources
*
**
*
Water treatment plant partially
treated
**
**
*
Water treatment plant fully treated
***
Distribution system
***
*** Normal Volume or First Choice ** Likely Volume
* Possible Volume
Summary of Contents for WSL25 Pro
Page 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Page 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Page 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Page 41: ...Section 5 HydroTest Photometer HT1000 ...
Page 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 141: ......
Page 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Page 152: ...NOTES ...