7.2 Single Point Calibration:
7.2.1 Press the MODE key until the meter shows the
indicator.
7.2.2 Rinse the conductivity electrode with distilled water, then rinse with a small amount of
calibration solution.
7.2.3 Press the CAL key; the meter enters the calibration mode.
7.2.4 Immerse the conductivity electrode in the calibration solution. The meter
automatically shows the detected calibration standard (e.g., 1413µS/cm).
7.2.5 Press ENTER key to confirm, “Calibration” indicator begins flashing.
7.2.6 Wait for the measured value to stabilise, the display shows “END”. Once calibration
is complete, the meter returns to measurement mode automatically.
7.3 Multi-Point Calibration:
7.3.1 Make sure that you have selected 2 to 5 points calibration in setup menu.
7.3.2 Repeat steps 1.2 to 1.5 above; when the first calibration point is completed, the
display will show “CAL2”. The meter prompts you to continue with the second calibration
point.