2.0 Incubator set-up
The incubator is integrated into the waterproof carry case with a docking station which
incorporates the rechargeable battery pack and power supply for the unit. An external charger is
also supplied which is connected to the port on the rear of the case.
The power pack is comprised of a 12V 15.6Ah Sealed
Non-Spillable
Lead acid battery. The battery
pack must be electrically isolated during transport on-board aircraft. To facilitate this, a switch has
been incorporated in the carry case which isolates the power supply.
Please note when transporting via aircraft the switch must be in the off position (depressed)
and secured such that it cannot become depressed during transit (covering the recessed plate
with rigid card should achieve this).
2.1 Docking Station
The Incubator is mounted on a docking station built into the carry case, which incorporates the
rechargeable battery pack and allows the incubator to be run from the batteries during
transportation. It is also possible to use the incubator on a bench independent of the dock if
desired, as long as it is connected to a 240 V AC outlet via the cables provided.
To remove the incubator from the docking station, lift the incubator vertically upwards, ensuring
not twist or tilt the incubator as this could result in damage to the connectors. To place the
incubator back onto the docking station, centre the incubator left to right on the docking station
with the front edge against the foam, then push down into position.
On/Off Switch
Charger
Connection
Summary of Contents for WSL25 Pro
Page 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Page 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Page 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Page 41: ...Section 5 HydroTest Photometer HT1000 ...
Page 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 141: ......
Page 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...