DISSOLVED OXYGEN ELECTRODE
•
Insert the 6-pin connector on the temperature probe into the corresponding connector on
the instrument, ensuring that the lugs are aligned. Once pushed firmly into place, tighten
the waterproof locking collar.
•
Once connected, DO NOT pull on the cables.
2.5 Inserting/Replacing the Batteries
The HC1000 is supplied pre-installed with the requisite battery.
The HC1000 requires 1 x 9V (PP3) battery.
To replace the battery, follow the below procedure:
1. Remove the battery cover from the rear of the meter.
2. Disconnect the snap connector and remove the battery.