Wait for a few seconds until the result is displayed in mg/l free Chlorine
.
Sample preparation (Total Chlorine):
•
Add
one HT15 DPD No.1 tablet
and
one HT16 DPD No. 3 tablet
straight from the foil to the
water sample.
•
Crush the tablets using a clean stirring rod, until no large pieces are visible.
•
Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
Analysis (Total Chlorine):
•
Place the vial in the sample chamber, making sure that the marks on the instrument and vial
are aligned.
•
Press the Test key
•
Wait for a reaction period of
2 minutes
•
After the reaction period, the measurement starts and the result is displayed in mg/l total
Chlorine
.
Sample preparation (Differentiated Determination):
•
Add
one HT15 DPD No.1 tablet
straight from the foil to the water sample.
•
Crush the tablet using a clean stirring rod, until no large pieces are visible.
•
Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
Analysis (Differentiated Determination):
•
Place the vial in the sample chamber, making sure that the marks on the instrument and vial
are aligned.
•
Press the Test key
Summary of Contents for WSL25 Pro
Page 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Page 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Page 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Page 41: ...Section 5 HydroTest Photometer HT1000 ...
Page 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Page 141: ......
Page 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Page 152: ...NOTES ...