7.3.3 Rinse the conductivity electrode with distilled water. Immerse the electrode in the next
calibration solution, the meter automatically shows the currently detected calibration
standard (e.g., 12.88mS/cm).
7.3.4 Press ENTER key to confirm, “Calibration” indicator begins flashing.
7.3.5 Wait for the measured value to stabilise, the display shows “CAL3”. The meter
prompts you to continue with third point calibration.
7.3.6 Repeat steps 2.3 to 2.4 above until the display shows “END”. Once calibration is
complete, the meter returns to measurement mode automatically.