6RL SERIES OPERATION MANUAL
SECTION 6: FRONT PANEL OPERATION
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
6RL Series Regenerative DC Load Operation Manual
Page 88 of 204
Some alarms are configurable:
Short Long
Description
Range
Indication
OVP
OverVoltage
Protection
Triggers an alarm if the DC Input voltage
reaches the defined threshold. The DC Input
will be switched off.
0 V...1.1*UNom
Display, analog &
digital interface
OCP
OverCurrent
Protection
Triggers an alarm if the DC Input current
reaches the defined threshold. The DC Input
will be switched off.
0 A...1.1*INom
Display, digital
interface
OPP
OverPower
Protection
Triggers an alarm if the DC Input power
reaches the defined threshold. The DC Input
will be switched off.
0 W...1.1*PNom
Display, digital
interface
The following alarms cannot be configured and are based on hardware:
Short
Long
Description
Indication
PF
Power Fail
AC supply over- or under voltage. Triggers an alarm if the AC
supply is out of specification or when the unit is cut from supply,
for example when switching it off with the power switch. The
DC Input will be switched off.
Display, analog &
digital interface
OT
OverTemp
etature
Triggers an alarm if the internal temperature reaches a certain
limit. The DC Input will be switched off.
Display, analog &
digital interface
MSP
Master-Slave
Protection
Triggers an alarm if the master unit loses contact to any slave unit.
The DC Input will be switched off. The alarm can be cleared by re-
initializing the M/S system.
Display, digital
interface
How to configure unit alarms:
1.
While the DC Input is switched off, tap the touch area on the main screen.
2.
On the right-hand side, tap the arrow to select “2. Protect”.
3.
Set the limits for the unit alarms relevant to your application if the default value of
110% is unsuitable.
Note: The set values can be entered using the decimal keypad. This will appear after
tapping the touch area “Direct input”.
The user also has the possibility of selecting whether an additional audible signal will be
sounded if an alarm or user defined event occurs.
How to configure the alarm sound (also see section 6.4.3, “MENU System Overview” on
page 61):
1.
While the DC Input is switched off tap the touch area
on the main
screen
2.
In the menu page, tap “HMI Settings”
3.
In the following menu page, tap “Alarm Sound”
4.
In the settings page tap on the symbol to either enable or disable the alarm sound
and confirm with