6RL SERIES OPERATION MANUAL
SECTION 7: REMOTE PROGRAMMING
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
6RL Series Regenerative DC Load Operation Manual
Page 149 of 204
Value range:
<NRf> = Min. resistance...max. resistance, according to technical
specs
Examples:
RES?
Absolute short form. Queries the currently set resistance
value.
SOUR:RESISTANCE˽10
Mixed form short/long. Sets 10 Ω.
SOURCE:RES˽MIN
Sets the resistance to the minimum defined for the
particular unit model.
7.12.7
Measurement Commands
Actual values, as returned by the measuring commands, are the DC output values as they
are present in the moment they are queried. They are not necessarily identical to the
corresponding set values. Your unit constantly measures the actual values and returns the
last snapshot when queried.
7.12.7.1
MEASure:[SCALar:]VOLTage[:DC]?
Queries the unit to return the last measured DC input resp. output voltage value in Volt.
Example:
MEAS:VOLT?
Absolute short form. Queries the actual voltage. A response,
which should be instantly coming from the unit, will return
a value between 0% and max. 125% of nominal unit voltage,
like for example “43.50 V”. The number of decimal places in
the returned value will be identical to the value format in
the unit display and varies from model to model.
7.12.7.2
MEASure:[SCALar:]CURRent[:DC]?
Queries the unit to return the last measured DC Input current value in Ampere.
Example:
MEASURE:CURRENT?
Queries the actual current only. A response, which should
be instantly coming from the unit, will return a value
between 0% and max. 125% of nominal unit cur- rent, for
example “100.1 A”. The number of decimal places in the
returned value will be identical to the value format in the
unit display and varies from model to model.
7.12.7.3
MEASure:[SCALar:]POWer[:DC]?
Queries the unit to return the last calculated DC input resp. output power value in Watts.
Example:
MEAS:POW?
Absolute short form. Queries the consumed (e-load) resp.
supplied power (PSU).