175.0020 - DECADE Elite service manual - rev 03
Trouble shooting
If during the upload procedure for any reason a power interruption occurs,
please repeat the upload procedure as described in this user chapter.
In such case do not repeat step 1 Export calibration setting, because
likely the detector will not contain any calibration settings anymore and
the valid original calibration file might be overwritten if step 1 is
repeated.
If something goes wrong during or after the firmware upgrade, the
DECADE Elite will not boot up correctly and will show the error message:
In that case repeat the upload procedure as described in this chapter. If
the error persists please contact your supplier.
Be careful with the calibration settings files. They are necessary to assure
that the detector operates properly after the software update. Take the fol-
lowing precautions.
•
Do not overwrite the first saved settings file during a second
attempt to export the settings file (rename file).
•
Always make a safety copy of the calibration file and save it
at a different location using the "save as" function. When
(factory) calibration settings are lost, please contact the sup-
plier for further instructions: [email protected]
74
E R R 2 0
C h e c k s u m e r
r o r