Chapter 5
Service menu
33
C H A P T E R 5
Service menu
This section describes the service menus accessible from keyboard/dis-
play. It is primarily used for trouble shooting in case a calibration parame-
ter is out of spec. For full calibration of the DECADE Elite/Lite the Elite Dia-
logue software is required, please refer to document
“171.0025 PM for
DECADE and ROXY
”.
The DECADE Elite service menu can be accessed by pushing the
“Enter”
button for more than 4 seconds, while in the main menu.
From the service mode main menu three sub menus can be accessed
which will be described in the next sections of this chapter:
•
SETT(tings) menu
•
TEST menu
•
ADJ(ustments) menu
In the top-right corner of the service mode main menu the active sensor
board is displayed
, to which the settings apply. The active sensor board
1
The service mode may only be accessed by authorized service
engineers. Incorrect settings may results into erratic operation of the
instrument.
A n t e c
M A I N
1
2
D E C A D E E L I T E
E C
D e t e c t o r
1 . 0 0
C O N F I G
> D C
P U L S E
S C A N
D I A G
> 4 seconds <ENTER> button
T h i s m o d e
i s
f o r
2D
q u a l
i
f
i e d s e r v
i c e p e r s o n n e l
o n l y
!
C o n t
i n u e ?
Y E S
N O
S e r v
i c e
m o d e
S E R V
1
47
P R E V
S E T T
T E S T
A D J
Summary of Contents for DECADE Elite
Page 3: ......
Page 14: ...14 175 0020 DECADE Elite service manual rev 03...
Page 22: ...22 175 0020 DECADE Elite service manual rev 03...
Page 46: ...46 175 0020 DECADE Elite service manual rev 03...
Page 68: ...68 175 0020 DECADE Elite service manual rev 03...
Page 72: ...72 175 0020 DECADE Elite service manual rev 03...