C H A P T E R 2
Required service tools
The following tools are necessary for servicing the DECADE Elite/Lite:
•
Screwdriver
•
Socket wrench 3/16”
•
Phillips screwdriver PH-1 and PH-2
Figure 1, Screwdriver, socket wrench and Phillips PH-1 screwdriver (from
left to right).
Calibrated temperature sensor
must have a tolerance at 45
o
C equal or
better than 0.1
o
C. Suggested model: Extech RTD 407907. Calibration cer-
tificate should not be older than 1 year.
Calibrated voltmeter
must have a resolution equal or better than 0.1 mV.
Suggested model: Fluke 12. Calibration certificate should not be older than
1 year.