CHAPTER 7
Replacing the power supply
67
Reinstallation of PSU frame
•
Remove the transparent plastic cover strip from the new PSU.
•
Fix all wiring to the PSU as shown in figure 21 and table IX de-
scribed on page 62.
Make sure that all wires are securely connected corresponding to the
wiring information on page 62. Incorrect wiring may lead to danger-
ous situation (electrocution) or damage to the instrument.
•
Reinstall the transparent plastic cover strip onto the PSU.
•
Place the PSU frame back into the PSU compartment in the bot-
tom panel of the detector.
•
Fix all 4 M3X8mm screws to secure the PSU frame.
•
Flip the detector back in its original position on the lab table
(standing on its instrument feet).
•
Reconnect the power cable into the mains inlet on the rear panel.
•
Connect an USB or LAN cable to the instrument if applicable.
•
Switch the instrument on by means of the mains switch on the rear
panel/
•
Check if the instrument powers up correctly. In case of a DECADE
Lite check if the instrument connects successfully with the Dia-
logue Elite or DataApex Clarity software.
In case the instrument does not start up correctly (Elite) or does not com-
municate over LAN/USB (Lite), check if all steps in this chapter are exe-
cuted correctly. If the problem persists please contact Antec for advice.