Chapter 6
Sensor board installation
47
C H A P T E R 6
Sensor board installation
This section describes the installation of an additional channel (sensor
board) in a DECADE Elite using the DECADE Elite additional channel kit
(p/n 175.0210). This kit enable you to upgrade your DECADE Elite to an
instrument with two or three flow cells.
Contents of the kit
In the table below the contents of p/n 175.0210 DECADE Elite additional
channel kit is listed:
P/n
Description
Qty
175.0210A
DECADE Elite mounting kit
1
175.0220
DECADE Elite/Lite internal cell cable
1
175.0604
DECADE Elite/Lite sensor board
1
250.0040
External dummy flow cell
1
250.0103K
Column clamp kit
1
171.0208
DECADE Elite/Lite cell cable
1
250.0102A
SenCell clamp kit
1
Carefully unpack the kit and inspect it for completeness and for possible
damage. Contact your supplier in case of damage or if not all marked
items on the checklist are included. Prior to shipment, the parts of your up-
grade kit have been inspected and tested to ensure the best possible per-
formance.
Preparation
As a precaution the following steps are taken:
1. Export the calibration settings to file using Dialogue software (File/Ex-
port settings). Name the file ~cal_data_nnn.ecd". Where nnn is any
new file counter. Save a second file with all device settings named f.e.
“settings_nnn.ecd”. Retrieving settings from file is only needed in case
anything unexpected goes wrong.
2. Ensure that the detector has the same firmware (FW) as the new
board. If not, follow the upgrade procedure (page 25) to the latest FW.
Summary of Contents for DECADE Elite
Page 3: ......
Page 14: ...14 175 0020 DECADE Elite service manual rev 03...
Page 22: ...22 175 0020 DECADE Elite service manual rev 03...
Page 46: ...46 175 0020 DECADE Elite service manual rev 03...
Page 68: ...68 175 0020 DECADE Elite service manual rev 03...
Page 72: ...72 175 0020 DECADE Elite service manual rev 03...