40
175.0020 - DECADE Elite service manual - rev 03
ZER.IE zero current for IE converters
Before adjusting the zero current make sure the system has been stabi-
lized for more than an hour with the cell off and T=35
o
C.
Procedure:
1. Select R using arrow and +/- buttons.
2. Set compensation to Ic = zero current using AUTO, it may be
necessary to use AUTO more than once.
3. Repeat procedure for each R
Table IV. Residual current specifications
R
Specified
1k
0.0 ± 0.1 μA
100k
0.0 ± 0.1 nA
1M
0.00 ± 0.01 nA
10M
0 ± 2 pA
100M
0 ± 2 pA
Table V. Parameters
Parameter Screen Description
Type
R
ZER.IE Select active resistor of IE converter using
+ or - button. Values are 100M, 10M, 1M,
100k, 1k Ohm
C
Ic
ZER.IE Residual current
C
ADC
ZER.IE Digital offset ADC
C
Offs
ZER.IE Analogue offset ADC
C
AUTO
ZER.IE Automatically compensate to zero current
F
UNDO
ZER.IE Restore previous compensation setting
F
RESET
ZER.IE Release any zero compensation
F
Explanation: Type S is status, F is function and C is control.
I c
= - 0 . 0 6 8 n A
Z E R .
I E
1
59
O f
f s
=
+ 1 0
R
=
1 M O h m
A D C
=
+ 1 8 8
2 5 > 3 0 ° C
P R E V
A U T O
U N D O
R E S E T
Cell cable must be disconnected all the time. If the cell cable is
connected ‘ERR 13 Disconnect flow cell’.
Summary of Contents for DECADE Elite
Page 3: ......
Page 14: ...14 175 0020 DECADE Elite service manual rev 03...
Page 22: ...22 175 0020 DECADE Elite service manual rev 03...
Page 46: ...46 175 0020 DECADE Elite service manual rev 03...
Page 68: ...68 175 0020 DECADE Elite service manual rev 03...
Page 72: ...72 175 0020 DECADE Elite service manual rev 03...