60
175.0020 - DECADE Elite service manual - rev 03
•
Mount the additional cell clamp and column clamps supplied in the
kit (in case the cells are used for detection in series, it is not nec-
essary to install the column kits).
•
Update the embedded software (FW) of the instrument if neces-
sary as described (page 25).
•
Let the instrument equilibrate with the oven set to 35°C for at least
half an hour.
•
Check and calibrate the instrument as described in Chapter 4 (ei-
ther via the SERVICE menu or using the PM procedure). Perform
a dummy cell test as described in the DECADE Elite IQ/OQ/PQ
manual (document 171.0023) to check if the instrument performs
within specification with respect to cell current and noise.