Spectrum Analyzer
2-33 Sweep Menu
Spectrum Analyzer MG
PN: 10580-00349 Rev. H
2-67
Gate Setup Menu
Key Sequence:
Shift
>
Sweep
(
3
) key
> Gated Sweep Setup
Gated Sweep:
Turns the Gated Sweep function On and Off. Gated Sweep
is not allowed in Zero Span mode. Refer to
Gate Source:
Selects the trigger source for the gated sweep.
External
: This setting designates the trigger source as an external
signal that is input via the instrument’s Ext Trigger In connector.
GPS (TD-LTE)
: The GPS trigger source is valid only on instrument
models with 20 MHz IF Bandwidth available:
Cell Master MT8212E/13E
Spectrum Master MS2712E/13E
Additionally, these instruments must be loaded with firmware V1.60
or later, and have these options installed:
Option 31 (GPS), and either of
Option 551 (TD-LTE/LTE-A RF Measurements), or
Option 883 (LTE/LTE-A FDD/TDD Measurements)
When GPS (TD-LTE) is selected, the gated sweep time is 10 ms.
IF Pwr
: This trigger type is available on model MS2720T only. When
choosing IF power as the trigger source, press the IF Trig Level key
(not shown) and use the rotary knob, arrow keys, or numeric keypad
to enter the trigger power level. The value ranges from -150 dBm to
+30 dBm.
Gate Polarity:
Selects the trigger edge to begin the gated sweep.
Gate Delay:
Sets the start of the gated sweep indicated by the left border
of the blue dashed rectangle shown in the bottom graph of
Gate Length:
Sets the length of the gate and is reflected on the zero span
graph by the width of the blue dashed rectangle.
Gate View Settings:
Opens the Gate View Settings menu. Allows a user
to independently change the RBW, VBW and sweep time of the zero span
or gate view (bottom graph).
Zero Span RBW:
Sets the resolution bandwidth of the zero span
graph.
Zero Span VBW:
Sets the video bandwidth of the zero span graph.
Zero Span Time:
Sets the sweep time of the zero span graph.
Back:
Returns to the Gate Setup menu.
Back:
and changes the Gated
Sweep Setup view back to the full screen spectrum view. The Gated Sweep
settings are retained and applied to the spectrum view.
Figure 2-52.
SPA Gated Sweep Menu
Gate Polarity
Rising Falling
Gate Setup
Back
Gate Source
External
Gate Delay
2
m
s
Gate Length
500
P
s
Gate View
Settings
Gate View Settings
Back
Zero Span RBW
100 kHz
Zero Span VBW
30 kHz
Zero Span Time
500
m
s
Gated Sweep
Off On
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com
Summary of Contents for Spectrum Master MS2720T
Page 2: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 5: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 22: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 241: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 264: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 265: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 558: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 800: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1027: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1033: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1034: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1035: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...