2-22 Spurious Emissions (Out-of-Band)
Spectrum Analyzer
2-32
PN: 10580-00349 Rev. H
Spectrum Analyzer MG
Loading an Existing Mask
1.
Press the
Shift
key followed by the
File
(
7
) key.
2.
Press
Recall
.
3.
Scroll through the list of files and select the desired spurious emission mask (file
extension .spe).
4.
Press
Enter
to load the selected mask into the Spurious Emission measurement.
Modifying an Existing Mask
To edit segments in an existing mask:
1.
Load the desired emission mask.
2.
Press
Segment Setup
to display the
Spurious Segments
submenu.
3.
Press the
Spurious Segment
submenu key and select the segment number to edit, which
is highlighted in red.
4.
Press the
Freq
main menu key followed by the
Start Freq
submenu key and enter a value
for the beginning point of a segment.
5.
Press the
Stop Freq
submenu key and enter a value for the end point of the segment.
6.
Return to the
Spurious Emissions
submenu by pressing the touch screen shortcut in the
upper right corner of the display, where the words
Spectrum Analyzer
are located.
You can also select the
Spurious Emissions
submenu by pressing the
Shift
key followed
by the
Measure
(
4
) key, then pressing the
Masks and C/I
submenu key followed by the
Spurious Emissions (Out-of-Band)
key.
7.
Press
Segment Setup
.
8.
Press
Segment Start Amplitude
and enter the desired amplitude for the first point of the
segment.
9.
Press
Segment Stop Amplitude
and enter the desired amplitude for the last point of the
segment.
10.
Press
Save Current Settings to Selected Segment
.
11.
Repeat
as necessary to edit more segments.
12.
Press
Save Mask
and enter the mask name in the
Filename
entry field. Use the existing
mask name if you wish to overwrite the file. If needed, select
Spurious Emission Mask
from the
Filetype
pull-down menu. Press
Enter
.
To add segments to an existing mask:
1.
Load the desired emission mask.
2.
Press
Segment Setup
to display the
Spurious Segments
submenu.
3.
Use the current segment number shown on the
Spurious Segment
submenu key or enter
the number of the segment to be used as a template for the new segment.
4.
Press
Add Current Settings As New Segment
.
5.
Press the
Freq
main menu key followed by the
Start Freq
submenu key and enter a value
for the beginning point of a segment.
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com
Summary of Contents for Spectrum Master MS2720T
Page 2: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 5: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 22: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 241: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 264: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 265: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 558: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 800: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1027: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1033: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1034: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...
Page 1035: ...ООО Техэнком Контрольно измерительные приборы и оборудование www tehencom com ...